TXCO-50 Series of carbon monoxide transmitters is a simple,
accurate and very economical way
to measure CO in the 0 to 50 ppm
levels. For special ranges please
consult Omega.
The TXCO-50 sensor uses a
special electro chemical cell to
measure the CO concentration.
The sensor meets the industrial
requirements for accuracy,
sensitivity and ease of use.
The transmitter comes fully
calibrated and requires no
calibration for two years. After
two years the TXCO-50 can be
calibrated with a CO gas. Typical
sensor life is up to 5000 ppm hrs.
Specifications Measuring Range: 0 to 50 ppm CO Special Ranges: Consult OMEGA Calibration: Using a known CO
calibration gas that is NIST traceable Accuracy:
±2% FSD T = constant
±5% FSD 0>T>50°C (122°F) Response Time: 90% FSD at
20°C (68°F) < 30 sec 90% Operating Temperature: 0 to 50°C
(32 to 122°F) Warm Up Time: 2 minutes Pressure: 0.1 to 1 bar Signal Output: 4 to 20 mA/DC
loop power Voltage: 24 Vdc Load: Type 470 Ω at 18 Vdc
maximum 700 Ω Housing: NEMA 4 (IP65) Dimensions: 80 L x 75 W x 55 mm D
(3.15 x 2.95 x 2.16") Enclosure: Painted aluminum Connection: ¼" tubing Weight: 1.0 kg (2.2 lb)
注: Comes complete with operator’s manual.